| Model Number |
30204E |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/19/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: bd insyte autoguard shielded iv catheter.The customer¿s report that the tubing ruptured during a reported power injection was confirmed.Visual inspection of the sets tubing area between the smartsite and male luer components to be torn.No other obvious damages or issues were observed.Functional testing was deemed unnecessary due to the obvious damage.The root cause was identified as use related as this is a non pressure-rated set and is not designed for power injector use.
|
| |
|
Event or Problem Description
|
|
It was reported that during a ct contrast power injection of the patient's abdomen and pelvis, the extension tubing ruptured and leaked.Prior to the ct, an ultrasound-guided iv was placed in right upper arm.There was no patient harm.
|
| |
|