| Catalog Number |
72202959 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/21/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a shoulder arthroscopy, the device was not working.The procedure was completed with a back up device with no patient injury reported.A delay greater than 30 minutes was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
An evaluation was performed by the supplier and could confirm the customer complaint for the device was not working.A visual inspection was performed and showed the scope to have distal tip and fiber damage, a bent outertube and a broken lens.This damage is caused by contact with another source.No manufacturing related defects were observed.
|
| |
|