| Model Number |
MMT-523LNAS |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/04/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump had motor error.Customer¿s blood glucose level was unknown.Customer reported that the drive support cap appeared to be normal.Customer was unable to clear the alarm.The customer was also advised that the insulin pump needed to be replaced and was advised to discontinue use of the insulin pump and to revert to the backup plan.The device will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
No motor error alarm or audio/beep anomaly noted during testing.Device passed rewind test, basic occlusion test, occlusion test, prime/a33 test, excessive no delivery test and displacement test.The motor was tested outside the device and passed.(b)(4).
|
| |
|