| Model Number |
PCB00 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Capsular Bag Tear (2639); Vitrectomy (2643)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Age/date of birth: unknown/ not provided.Sex/gender: unknown/ not provided.Date of event: date unknown/ not provided.If implanted; give date: lens was not implanted.If explanted; give date: lens was not implanted, therefor not explanted.(b)(6).The intraocular lens (iol) is not returning for evaluation as it was discarded; therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, labeling, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specification.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was initially reported that the customer had issue with a preloaded device pcb00.Through follow-up we learned that the posterior capsule ruptured as the lens was introduced into the eye requiring a vitrectomy to be performed.The lens was removed again from the eye and a sulcus lens was implanted instead.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
|
| |
|
Search Alerts/Recalls
|