Model Number 7010
Device Problem Material Deformation (2976)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/06/2019
Event Type  malfunction  
Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of a quantum maverick balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There were numerous kinks.There was a complete separation at 69.6cm distal of the strain relief.There was blood in the guidewire lumen.The balloon was tightly folded.Microscopic inspection revealed tip damage.Inspection of the remainder of the device presented no other damage or irregularities.
 
Event Description
Reportable based on device analysis completed on 20aug2019.It was reported that shaft kink occurred.A 2.0mm x 12mm quantum maverick balloon was advanced for dilatation.However, it was noted that the balloon delivery shaft was kinked.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.However, returned device analysis revealed shaft detached/separated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUANTUM MAVERICK
Type of Device
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key9005731
MDR Text Key157887174
Report Number2134265-2019-11104
Device Sequence Number1
Product Code LOX
UDI-Device Identifier08714729392637
UDI-Public08714729392637
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P860019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 09/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/12/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/02/2021
Device Model Number7010
Device Catalogue Number7010
Device Lot Number0022744294
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/22/2019
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/20/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/03/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age55 YR
Patient Weight58
-
-