• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. PRONOVA* BLUE MONO; SUTURE, NON ABSORBABLE, SYNTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. PRONOVA* BLUE MONO; SUTURE, NON ABSORBABLE, SYNTHETIC Back to Search Results
Catalog Number D10114
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/16/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4.A manufacturing record evaluation was performed for the finished device batch mmh592- d10114 and no non-conformance's were identified.Attempts are being made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The following additional information was obtained: can you confirm and provide the specific number of devices (total qty actually involved with reported issue) that failed during use on same patient/ single procedure ? t is reported 3 devices.Events reported via mw # 10968-2019-87143, 2210968-2019-87144, 2210968-2019-87145.
 
Event or Problem Description
It was reported that the patient underwent an unknown cardiovascular surgery on (b)(6) 2019 and suture was used.During the procedure, the needle was detached from the suture during continuous suturing.It was not a control release needle.Another product in another lot number was used with no problem.There were no adverse consequences to the patient.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).Device evaluation summary: it was received for analysis eighty unopened samples of product code d10114, lot mmh592.The complaint sample was not received for evaluation.During the visual inspection of thirteen samples, no defects were found on the packages.The samples were opened, and the swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along of the strands and no defects were observed.A functional test was performed and the pull force were above the minimum requirements.The manufacturing records were reviewed, and the manufacturing/ packaging criteria were met prior to the release of this batch.Per the condition of the samples, no performance pull off suture needle was found, and the tested samples met the finished goods requirements.Representative samples returned to support investigation of 3 device issues captured in reports 2210968-2019-87143, 2210968-2019-87144 and 2210968-2019-87145.Analysis results have been included in follow up reports for each.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
SUTURE, NON ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876 0151
MDR Report Key9007656
Report Number2210968-2019-87143
Device Sequence Number16746767
Product Code GAW
UDI-Device Identifier10705031203396
UDI-Public10705031203396
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K001625
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2023
Device Catalogue NumberD10114
Device Lot NumberMMH592
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/15/2019
Supplement Date Received by Manufacturer09/13/2019
Initial Report FDA Received Date09/12/2019
Supplement Report FDA Received Date10/10/2019
Patient Sequence Number1
-
-