Lot Number 084655
Medical Device Problem Code Device Contamination with Body Fluid (2317)
Health Effect - Clinical Code Deposits (1809)
Date of Event 05/21/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Based on the initial report (documentation analysis, complaint history and investigation of the returned lens) and the assessment by lenstec's medical monitor, lenstec concludes that there is no evidence to suggest that any aspect of the lens was responsible for this incident.Microscopic examination of the lens indicated that the bubbles seen were deemed superficial, stained for calcium and did not extend through the matrix of the lens.According to neuhann et al, non-lens factors are the likely cause of these bubbles seen.However, lenstec is unable to determine a definitive conclusion as to the cause.Additionally, lenstec can confirm that the lens, its design and the manufacturing process are not at fault due to the extensive testing that we have performed on this model.Furthermore, there have never been any confirmed lens-related cases of clouding, discoloration or opacification for our hema lenses.
 
Event or Problem Description
Lenstec received an email stating "lens needed to be explanted on the 21/5 due to glistening.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOFTEC I
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
LENSTEC BARBADOS INC.
airport commercial centre
pilgrim road
christ church,
BB 
MDR Report Key9008597
Report Number9613160-2019-00023
Device Sequence Number1890941
Product Code HQL
UDI-Device Identifier00844369029766
UDI-Public00844369029766
Combination Product (Y/N)N
PMA/510(K) Number
P090022
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2019
Reporter Type Manufacturer
Report Source foreign
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/15/2013
Device Lot Number084655
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/02/2019
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/29/2019
Initial Report FDA Received Date09/12/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-