• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EVOLUTION DUODENAL CONTROLLED-RELEASE STENT - UNCOVERED; MUM STENT, METALIC EXPANDABLE, DUODENAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EVOLUTION DUODENAL CONTROLLED-RELEASE STENT - UNCOVERED; MUM STENT, METALIC EXPANDABLE, DUODENAL Back to Search Results
Model Number G48027
Medical Device Problem Code Activation, Positioning or Separation Problem (2906)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Date of Event 08/15/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
510(k) number: k163468.Investigation is still pending, a follow up mdr will be submitted to include the investigation conclusions.
 
Event or Problem Description
Patient's stenosis is 8 cm long.User cannot see the stent under x-ray image during stent deployment, and there was a sound like slam while the marker at 1/3 position, but user didn't stop deployment after the sound and pulled the safety wire out.The stent cannot be saw during the whole deployment because of the contrast accumulation, and the yellow marker cannot be saw either because it was covered by excrement.User complete the deployment of stent by experience until marker reached terminal point and the trigger cannot be pulled.User retracted the device from patient afterward and found out the stent was still inside sheath.User changed another same device to complete the procedure.
 
Additional Manufacturer Narrative
510(k) number: k163468 this report is being submitted as a cancellation report.From the initial information received, this event was cautiously made reportable based on the precedence in place for this device of 'flexor kinked/stretched/broken/compressed'.However, when the returned device was received it was confirmed during the lab evaluation that this malfunction did not occur.The following failure has been confirmed: "shuttle assembly does not remain intact, stent cannot be deployed.Replacement device required".Overall risk assessed as category iii (low).No reporting malfunction precedence exists for this complaint event for this product family.Low risk of failure mode indicates no potential for serious injury if the malfunction were to recur.
 
Event or Problem Description
Patient's stenosis is 8 cm long.User cannot see the stent under x-ray image during stent deployment, and there was a sound like slam while the marker at 1/3 position, but user didn't stop deployment after the sound and pulled the safety wire out.The stent cannot be saw during the whole deployment because of the contrast accumulation, and the yellow marker cannot be saw either because it was covered by excrement.User complete the deployment of stent by experience until marker reached terminal point and the trigger cannot be pulled.User retracted the device from patient afterward and found out the stent was still inside sheath.User changed another same device to complete the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOLUTION DUODENAL CONTROLLED-RELEASE STENT - UNCOVERED
Common Device Name
MUM STENT, METALIC EXPANDABLE, DUODENAL
MDR Report Key9009407
Report Number3001845648-2019-00470
Device Sequence Number2332537
Product Code MUM
UDI-Device Identifier10827002480275
UDI-Public(01)10827002480275(17)200822(10)C1532216
Combination Product (Y/N)N
Initial Reporter CountryCN
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG48027
Device Catalogue NumberEVO-22-27-12-D
Device Lot NumberC1532216
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/03/2019
Was the Report Sent to FDA? No
Distributor Facility Aware Date08/15/2019
Event Location Hospital
Initial Date Received by Manufacturer 08/16/2019
Supplement Date Received by Manufacturer08/16/2019
Initial Report FDA Received Date09/12/2019
Supplement Report FDA Received Date10/10/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age48 YR
-
-