| Model Number |
VA8074 |
| Medical Device Problem Codes |
Material Rupture (1546); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Injury (2348)
|
| Date of Event |
08/19/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.Medical device - expiry date: 05/2022.
|
| |
|
Event or Problem Description
|
|
It was reported that during an angioplasty procedure of a venous outflow graft, the pta balloon allegedly ruptured and had detached material.It was further reported that the patient required additional hospitalization to retrieve the detached material.The patient is reportedly stable post procedure.
|
| |
|
Event or Problem Description
|
|
It was reported that during an angioplasty procedure of a venous outflow graft, the pta balloon allegedly ruptured and had detached material.It was further reported that the patient required additional hospitalization to retrieve the detached material.The patient is reportedly stable post procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: the device history records were reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.The device was returned for evaluation; device evaluation confirmed detachment of the device or a device component, but was inconclusive for material rupture as the balloon portion of the device was not returned.The definitive root cause for the reported issues could not be determined based upon available information.It is unknown whether patient or procedural issues contributed to the event.The review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.H10: d4: (expiry date: (b)(6) 2022.
|
| |
|