Model Number 24970A |
Device Problems
Pacing Problem (1439); Impedance Problem (2950)
|
Patient Problem
Cardiac Arrest (1762)
|
Event Date 08/16/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that after the physician requested a lead impedance check and this check was initiated using the mobile programmer application the pacing was lost, causing asystole.Upon ending the test pacing was resumed.The programmer was changed out and a different model was brought in to recheck the test.It was noted that the test was unable to be performed until the pace and sense polarities were programmed but that then the test was able to be performed without any change in heart rhythm.No further patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|