ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ACCENT DR RF PACEMAKER; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM2210
Device Problem Pacemaker Found in Back-Up Mode (1440)
Patient Problem Heart Failure (2206)
Event Date 08/19/2019
Event Type  malfunction  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that the patient presented to the emergency department stating was not feeling well.Upon interrogation, it was noted that the pacemaker was in backup bvvi, resulting from an unsuccessful firmware update.The loss of av synchrony, may have contributed to the patient¿s acute heart failure symptoms.The device was successfully restored.The patient was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCENT DR RF PACEMAKER
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
elizabeth boltz
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key9017388
MDR Text Key158109764
Report Number2017865-2019-13247
Device Sequence Number1
Product Code LWP
UDI-Device Identifier05414734503617
UDI-Public05414734503617
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 09/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/12/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2012
Device Model NumberPM2210
Device Lot Number3278384
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/29/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/22/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-