| Model Number |
MMT-523RNAS |
| Medical Device Problem Codes |
Failure to Cycle (1142); Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
Hypoglycemia (1912)
|
| Date of Event |
09/07/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that they experienced low blood glucose and insulin pump alarmed compromised force sensor system.The customer¿s blood glucose level was 35 mg/dl.The customer also reported that the insulin was squirting during manual prime process.The customer stated that the drive support cap was normal.The customer was advised that the insulin pump needed to be replaced and was advised to discontinue use of the insulin pump and to revert to the backup plan.The insulin pump will not be returned for analysis.
|
| |
|