COOK INC UNKNOWN; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Catalog Number UNKNOWN
Device Problems Structural Problem (2506); Material Twisted/Bent (2981); Insufficient Information (3190); Appropriate Term/Code Not Available (3191)
Patient Problems Vessel Or Plaque, Device Embedded In (1204); Internal Organ Perforation (1987); Pain (1994); Perforation of Vessels (2135); Abdominal Cramps (2543); No Information (3190)
Event Type  Injury  
Event Description
It is alleged that the patient received a gunther tulip on (b)(6) 2008.It is alleged that the patient was injured without further explanation.Hospital and medical records have been requested but not yet provided.
 
Event Description
Patient allegedly received an implant on (b)(6) 2008 via right common femoral vein due to motor vehicle accident.Patient is alleging tilt and vena cava perforation.The patient further alleges ¿pain and cramping in lower stomach, constipation and back pain¿ as well as limited physical activity.Per a ct (computed tomography) scan of the abdomen dated (b)(6) 2019, ¿aberrant placement ivc filter with 25% degree tilting.Bent 1 o'clock strut.Perforated struts at 12 and 2 o'clock.¿.
 
Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.H6 device code(s): appropriate term/code not available (3191) was selected for the alleged device tilt.H6 device code(s): appropriate term/code not available (3191) was selected for the alleged perforation.Investigation: investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.The following allegations have been investigated.Vena cava (vc) perforation, bent, tilt, pain, cramping, constipation and limited physical activity.Filter interacts with ivc wall, e.G.Penetration perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration.Filter tilt has been reported.Potential causes may include filter placement in ivcs with diameters larger than those specified in these instructions for use; improper deployment; manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter); and (or) a failed retrieval attempt.Excessive filter tilt may contribute to difficult or failed retrieval; vena cava wall penetration/perforation; and (or) result in loss of filter efficiency.Potential adverse events that may occur include, but are not limited to, the following: unacceptable filter tilt.Unknown if the reported bent, pain, cramping, constipation, limited physical activity are directly related to the filter and unable to identify a corresponding failure mode at this point in time.Catalog and lot numbers are unknown, however, the device is manufactured and inspected according to current controls.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
The patient alleges organ perforation and ivc wall/bowel/aorta embedment.On (b)(6) 2019, per a report from computed tomography 2; ¿the hook of the cook gunther tulip retrievable ivc filter is at the l1-2 interspace.The ivc filter is severely tilted posteriorly and laterally and the hook is embedded within the ivc wall.One (1) anterior strut perforates the ivc wall 13mm and is embedded within the bowel.One (1) medial strut perforates the ivc wall 15mm and is embedded within the aorta.Thrombus within the ivc or filter cannot be evaluated on this non contrast study.No fracture fragments are identified.¿.
 
Manufacturer Narrative
Investigation: the following allegations have been investigated: organ perforation, ivc wall/bowel/aorta embedment, back pain.Investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.The additional information regarding organ perforation, ivc wall/bowel/aorta embedment does not change the previous investigation results for vena cava perforation.Unknown if the reported back pain is directly related to the filter and unable to identify a corresponding failure mode at this point in time.Catalog number and lot number are unknown, however, the device is manufactured and inspected according to current controls.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key9022101
MDR Text Key158108483
Report Number1820334-2019-02294
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup,Followup
Report Date 02/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 08/26/2019
Initial Date FDA Received09/13/2019
Supplement Dates Manufacturer Received10/25/2019
01/19/2021
Supplement Dates FDA Received11/13/2019
02/01/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Life Threatening; Other;
Patient Weight90
-
-