| Catalog Number |
257002200 |
| Medical Device Problem Codes |
Naturally Worn (2988); Scratched Material (3020)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/22/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: examination of the returned instrument confirmed the complaint.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Visual analysis found nicks at the tip of the attachment end.The flat areas on the attachment end are stripped / worn.The fluted area of the reamer also exhibits nicks with missing material.The root cause is attributed to wear out.For the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
We were informed that the summit tapered reamer size 4/5 is damaged at the connecting end.There was no surgical delay.The surgery was completed without delay.There looked to be slight damage to the end of the reamer where it connects to the t-handle or reamer attachment.Lot # for the item is a0905.
|
| |
|