Product complaint # (b)(4).Udi: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Upon visual inspection, it is observed that the dual threaded screw got broken at head part near the cap and separated into three pieces (cap, head and shaft).The rest of the device shows normal wear which would not contribute to the complaint condition.A review of the device history record was conducted.No issues were identified during the manufacturing and release of this product that could have contributed to the problem reported by the customer.A visual inspection, and document/specification review were performed as part of this investigation.The dual threaded screw got broken at the head part.Thus, the complaint is confirmed.While a definitive root cause could not be identified, it is possible that the device might have encountered unintended forces.No product design or manufacturing issues were identified, and no new malfunctions were identified either.Based upon these results, no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.
|