Model Number 71674540
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/14/2019
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the customer attempted to use the combined integrated lag and compression screw method into the neck of femur, however once the compression starter drill was used it appeared to catch on the jig/nail causing significant drill damage and insufficient drill depth for the compression screw.As an alternative the decision was made to use a single lag screw instead which again came into contact with the jig/nail and caused significant damage to the lag screw.No more information provided.
 
Additional Manufacturer Narrative
The devices, used in treatment, were not returned for evaluation.The photo provided confirmed the tip is broken on the drill.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.The devices are reusable instruments that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Without the actual product involved, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERTAN LAG SCREW DRILL
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key9038252
Report Number1020279-2019-03349
Device Sequence Number1743398
Product Code HSB
UDI-Device Identifier00885556363348
UDI-Public00885556363348
Combination Product (Y/N)N
PMA/510(K) Number
K122170
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71674540
Device Catalogue Number71674540
Device Lot NumberUNKNOWN/UNREADABLE
Initial Date Received by Manufacturer 08/14/2019
Supplement Date Received by Manufacturer02/06/2020
Initial Report FDA Received Date09/13/2019
Supplement Report FDA Received Date04/17/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-