| Model Number |
ES28202400 |
| Medical Device Problem Codes |
Inadequacy of Device Shape and/or Size (1583); Migration (4003)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
07/31/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Without the benefit of examination and testing, coloplast is precluded from commenting on the condition of the device or the cause of the occurrence.Should additional facts prompt us to alter or supplement any information or conclusions contained in the original mdr or in any prior supplemental reports, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
According to the available information, only the rear tip extenders were replaced with a different size.The device appeared to inflate/deflate without difficulty; however the right side cylinder had crossover (distal).
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up mdr is created to document the corrected device information and implant date, and the conclusion of the investigation.The device was not received for evaluation.As examination of the device may not conclusively confirm or disprove the report of crossover quality accepts the physician's observations as to the reason for surgical intervention.A review of the device history record confirmed the devices from this lot met all specifications prior to release.A review of the complaint history database revealed no trends for this lot.Review of nonconforming reports revealed no nonconformances for this lot.No capas are associated with this lot.
|
| |
|
Search Alerts/Recalls
|