| Model Number |
TCM-375 |
| Medical Device Problem Code |
Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
07/19/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture, baker grade iii.This is a known potential adverse event addressed in the product labeling.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side capsular contracture, baker grade iii.Patient additionally reported capsular contracture.Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Allergan did not submit this mdr within 30 days of becoming aware.Recent stimulated reporting related to 2011068-7/2/19-001-r has increased complaint and mdr volume.Allergan is implementing a plan to address the increased volumes. device evaluation: visual analysis of the returned device identified: deformation, wear abrasion, cloudy and creases.A weight test of the device was verified and the device was within specification.Based on the device analysis the final assessment is: no issues found related with the manufacturing.
|
| |
|
Event or Problem Description
|
|
Healthcare professional additionally reported reason for removal as "textured" and also reported a right side double capsule.Device has been explanted.
|
| |
|
Search Alerts/Recalls
|