| Model Number |
MMT-523RNAL |
| Medical Device Problem Codes |
Device Difficult to Program or Calibrate (1496); Visual Prompts will not Clear (2281)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/08/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Customer reported via phone call that the insulin pump received button error alert and frozen display.The customer¿s blood glucose was 138 mg/dl.The customer stated that the they observe time clock for one minute to verify if time did not advanced.Customer stated that alarm did not recur.The customer was advised that the insulin pump needed to be replaced and was advised to discontinue use of the insulin pump and to revert to the backup plan per health care professional instructions.The device will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Insulin pump received with no (act, up and escape) button response due to corroded keypad traces.No button error alarm and no frozen display anomaly during testing noted.Unable to perform all functional testing including the displacement, operating currents, a21 error test, rewind, basic occlusion, occlusion, prime a33 and excessive no delivery test due to buttons not responding.
|
| |
|