|
The axium pushwire was returned for analysis without the microcatheter or the accessories devices.The axium pushwire was decontaminated.The axium pushwire was found to be inverted within the introducer sheath with the wavelock at the distal end.The axium pushwire was found bent and kinked within the introducer sheath.No damage was found with the introducer sheath.The coin was found against the lumen stop.The shield coil was present but stretched with the implant coil broken off and not returned.The shield coil was removed to reveal that the detach element was broken between the ball and loop causing the implant coil to separate.Under microscope, the jacket was removed, and the coin and release wire were pulled back, releasing the detach element ball.No other anomalies were observed.Based on the analysis performed, the report of ¿coil stretch¿ could not be confirmed as the implant coil was not returned.Possible causes for coil stretch are advancing or retracting the coil against the reported resistance or by use of an improper hubbing technique (over tightening of the rhv/sheath not firmly seated into hub).The distal section of the pushwire was found bent/kinked.The potential cause for this failure is advancing the coil against the reported resistance.The implant coil was found broken at the detach element, but the cause could not be determined.Possible causes for ¿coil separation/break¿ are: coil is not retracted in a one-to-one motion with the implant pusher during repositioning, pushwire rotation or, user advances the coil against resistance.As the axium pushwire was found to be inverted within the introducer sheath, it is also possible that resistance would occur when attempting to push the implant coil out from this portion of the introducer sheath as this portion of the introducer sheath is designed to create resistance with the pushwire portion of the axium coil.It is likely that the customer inadvertently inverted the axium coil following hydration or inspection.Pushing/pulling the axium implant coil through the wave-lock portion of the introducer sheath would potentially cause the implant coil to become damaged.There is no malfunction of the pushwire that would contribute to the issue and pulling back the coin during analysis released the detach element as intended.Per the detachable coil and i.D.(instant detacher) instructions for use (ifu): warning: "if the coil does not move with a one-to-one motion, or repositioning is difficult, the coil has been stretched and could possibly break.Gently remove and discard both the catheter and coil." if information is provided in the future, a supplemental report will be issued.
|