SMITH & NEPHEW, INC. MOTOR DRIVE UNIT HAND CNTRL PWRMX EL; SAW, POWERED, AND ACCESSORIES Back to Search Results
Catalog Number 72200616
Device Problem Mechanical Problem (1384)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/22/2019
Event Type  malfunction  
Event Description
It was reported that during a knee procedure, 2 buttons of the device were going on and off.Backup device was available to complete the procedure.No delay and no patient injuries were reported.
 
Manufacturer Narrative
The device was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed and no issues were noted.Mdu would start and stop running each time the housing was slightly shaken.Cause of intermittent start/stop is a corroded right button assembly.Mdu also fails intermittently for hand piece error when plugged into the control unit.The complaint was confirmed and the root cause has been determined to be corrosion of the right button assembly.Factors that could have contributed to the event include a buildup of corrosion/debris around the button assembly from cleaning chemical fluid ingression over time.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOTOR DRIVE UNIT HAND CNTRL PWRMX EL
Type of Device
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
holly topping
7000 w william cannon drive
austin, TX 78735
5123913906
MDR Report Key9070177
MDR Text Key164298225
Report Number1643264-2019-00640
Device Sequence Number1
Product Code HAB
UDI-Device Identifier03596010555472
UDI-Public03596010555472
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72200616
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/09/2019
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/22/2019
Initial Date FDA Received09/16/2019
Supplement Dates Manufacturer Received10/16/2019
Supplement Dates FDA Received10/20/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-