COOK INC COOK SILICONE BALLOON HYSTEROSALPINGOGRAPHY INJECTION CATHETER; HES INSUFFLATOR-CARBON DIOXIDE, URETOTUBAL (AND ACCESSORIES) Back to Search Results
Model Number G17147
Medical Device Problem Code Difficult to Remove (1528)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown or unavailable.Occupation: account manager.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event or Problem Description
During an unspecified procedure, four cook silicone balloon hysterosalpingography injection catheter were used.The complainant stated the balloon would inflate, but that when the procedure was finished it would not deflate.No adverse effects to the patient have been reported due to this event.Additional information has been requested and a follow-up report will be submitted when/if the requested information is received.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unchanged, unknown or unavailable.Investigation - evaluation.Reviews of complaint history, device history record, manufacturing instructions, quality control data, and the instructions for use(ifu) were conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record found no non-conformances related to the reported failure mode.A review of complaint history records shows no other complaints associated with the complaint device lot.Because there are no related non-conformances, adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other lot related complaints that have been received from the field, it was concluded that there is no evidence that nonconforming product exists in house or in field.The instructions for use (ifu), provides the following information to the user related to the reported failure mode: warnings: "always inflate the balloon with a sterile liquid.Never inflate with air, carbon dioxide or any other gas." how supplied: "upon removal from the package, inspect the product to ensure no damage has occurred." the complainant did not return the complaint device for investigation.At the time of the complaint investigation, all products in the complaint device lot have been distributed to customers, so no representative product is available from the lot for evaluation.A definitive cause could not be determined from the available information.Per the quality engineering risk assessment, the risk analysis for this failure mode was reviewed, and it was determined that no additional risk mitigating activity is required.Cook medical will continue to monitor this device via the complaints database for similar complaints.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No new patient or event information since last report was sent on (b)(6) 2019.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOK SILICONE BALLOON HYSTEROSALPINGOGRAPHY INJECTION CATHETER
Common Device Name
HES INSUFFLATOR-CARBON DIOXIDE, URETOTUBAL (AND ACCESSORIES)
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key9072300
Report Number1820334-2019-02339
Device Sequence Number3042202
Product Code HES
UDI-Device Identifier00827002171473
UDI-Public(01)00827002171473(17)220517(10)9748946
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K891290
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/17/2022
Device Model NumberG17147
Device Catalogue NumberJ-CHSG-503000
Device Lot Number9748946
Initial Date Received by Manufacturer 09/03/2019
Supplement Date Received by Manufacturer10/14/2019
Initial Report FDA Received Date09/16/2019
Supplement Report FDA Received Date10/14/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-