Model Number 27MECJ-502
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/15/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Further information regarding this event has been requested.The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
 
Event or Problem Description
On (b)(6) 2017, a 27mm masters series valve was implanted.On (b)(6) 2019, the valve was explanted due to leaflet obstruction and the valve was exchanged with a 27mm epic stented porcine valve.The patient has been discharged and no patient consequences were reported.
 
Additional Manufacturer Narrative
The reported event of a "stuck" valve could not be confirmed.The investigation found that both leaflets opened and closed completely and easily.Pannus was present on the remains of the sewing cuff but did not impinge or obstruct the leaflets.No acute inflammation or significant calcifications were present.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications at the time of commercialization.The cause of the reported event could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM MASTERS SERIES VALVE EXPANDED CUFF
Common Device Name
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ST. JUDE MEDICAL PUERTO RICO, B.V.
lot 21
caguas west industrial park
caguas, puerto rico 00726 -099
MDR Report Key9075774
Report Number3003681312-2019-00006
Device Sequence Number2475542
Product Code LWQ
UDI-Device Identifier05414734006453
UDI-Public05414734006453
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
P810002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/04/2022
Device Model Number27MECJ-502
Device Catalogue Number27MECJ-502
Device Lot Number5798388
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/10/2019
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 08/21/2019
Supplement Date Received by Manufacturer10/24/2019
Initial Report FDA Received Date09/17/2019
Supplement Report FDA Received Date10/29/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age54 YR
-
-