• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - BATTERY; VENTRICULAR (ASSIST) BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - BATTERY; VENTRICULAR (ASSIST) BYPASS Back to Search Results
Model Number 1650DE
Medical Device Problem Codes Battery Problem (2885); Connection Problem (2900)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/24/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device is used for treatment not diagnosis.The ventricular assist system is indicated for use as a bridge to cardiac transplantation and destination therapy in patients who are at risk of death from refractory end-stage left ventricular heart failure.The system is designed for in-hospital and out-of-hospital settings.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the battery had a poor mechanical connection and power disconnect alarms.After routine log file review, it was indicated that the battery had a communication error.The battery was removed from service.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the battery was not returned for evaluation.As a result, the reported poor mechanical connection event could not be confirmed.Log file analysis revealed that the controller in use during the reported event contained a feature that records whether a power source experienced a communication error or a disconnection within each 15 minute interval.Data log files revealed 23 instances involving (b)(6) where the battery¿s relative state of charge (rsoc) was between 101-201, which are indicative of communication errors.Applicable risk documentation and experience with events of similar circumstances were considered; a possible root cause of the reported poor mechanical connection event can be attributed, but not limited, to connector damage.Possible root causes of the reported communication error and power disconnect alarms can be attributed to momentary disconnections on the communication pins of the controller, the controller not receiving responses from the battery, and/or due to the packet error checking method detecting bit errors.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTWARE VENTRICULAR ASSIST SYSTEM - BATTERY
Common Device Name
VENTRICULAR (ASSIST) BYPASS
Manufacturer (Section D)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
MDR Report Key9077742
Report Number3007042319-2019-10926
Device Sequence Number19445828
Product Code DSQ
UDI-Device Identifier00888707000376
UDI-Public00888707000376
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P100047
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 10/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date12/31/2017
Device Model Number1650DE
Device Catalogue Number1650DE
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/26/2019
Supplement Date Received by Manufacturer10/18/2019
Initial Report FDA Received Date09/17/2019
Supplement Report FDA Received Date10/18/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
1103 VAD; 1103 VAD
Patient Age63 YR
-
-