MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. VISIA AF MRI S VR SURESCAN; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. VISIA AF MRI S VR SURESCAN; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER Back to Search Results
Model Number DVFC3D4
Medical Device Problem Code Premature Discharge of Battery (1057)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/23/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the implantable cardioverter defibrillator (icd) triggered end of service (eos) and premature battery depletion was suspected.The icd was explanted and replaced.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the device was returned and analyzed.The returned device indicated shorting/low impedance in the battery.Hybrid analysis revealed that the device loss of telemetry and all functions were due to a severely depleted battery.The device was fully functional and operated under normal current drain when powered with an external supply.There was no evidence of a high current drain condition, and no hybrid anomalies were observed.Battery analysis revealed that lithium plating was found to initiate from the case, extended under the head space insulator (hsi) and contacted the cathode tab.The tab is located in segment 3 of the coil.Based on this analysis, the premature battery depletion was the result of an internal short from the plated lithium material bridging the case (negative polarity) and cathode tab (positive polarity).If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISIA AF MRI S VR SURESCAN
Common Device Name
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
MDR Report Key9078030
Report Number3004209178-2019-17803
Device Sequence Number2451620
Product Code LWS
UDI-Device Identifier00643169566446
UDI-Public00643169566446
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P980016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date05/14/2018
Device Model NumberDVFC3D4
Device Catalogue NumberDVFC3D4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/09/2019
Initial Date Received by Manufacturer 08/29/2019
Supplement Date Received by Manufacturer02/28/2020
03/27/2021
Initial Report FDA Received Date09/17/2019
Supplement Report FDA Received Date03/04/2020
04/19/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1305-2021
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
6935M LEAD; 6935M LEAD
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age57 YR
-
-