Model Number 3660
Device Problem Application Program Freezes, Becomes Nonfunctional (4031)
Patient Problem Inadequate Pain Relief (2388)
Event Date 08/27/2019
Event Type  malfunction  
Manufacturer Narrative
The device is included in the neuromodulation implantable pulse generator (ipg) inoperable when exposed to monopolar electrosurgery advisory notice issued by abbott on 02 june 2017.The results / method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
It was reported that patient¿s ipg failed to established communication with external devices after a cervical surgery.Reportedly, the ipg was not set into surgery mode during the procedure.Field representative was able to recover connection after troubleshooting.Issue has been resolved.
 
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Actions have been taken to prevent reoccurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROCLAIM 5 ELITE
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer Contact
andrea deitz
6901 preston road
plano, TX 75024
9723098000
MDR Report Key9079744
MDR Text Key159038301
Report Number3006705815-2019-03483
Device Sequence Number1
Product Code LGW
UDI-Device Identifier05415067020192
UDI-Public05415067020192
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/17/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/13/2020
Device Model Number3660
Device Lot NumberA000060321
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/26/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/14/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1627487/06/02/2017/001-C
Patient Sequence Number1
-
-