| Catalog Number |
04.01.0161 |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/21/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Batch review performed on 02.Sep.2019.Lot 184569: 303 items manufactured and released on 15-jan-2019.Expiration date: 2023-12-05.No anomalies found related to the problem.To date, 65 items of the same lot have been already sold without any other similar reported event.
|
| |
|
Event or Problem Description
|
|
During the primary reverse shoulder surgery and while inserting the 30mm glenoid polyaxial locking screw, the tabs on the screw head bent causing the screw not to seat or lock.The surgeon removed the screw and was able to successfully implant, seat and lock the new screw.There was no delay in the case and the surgery was completed successfully.
|
| |
|
Search Alerts/Recalls
|