MAUDE Adverse Event Report: ENCORE MEDICAL L.P. BIPOLAR HIP; SHELL, BIPOLAR, MODULAR, ENCORE, SIZE45

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. BIPOLAR HIP; SHELL, BIPOLAR, MODULAR, ENCORE, SIZE45 Back to Search Results
Model Number 412-02-045
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 08/19/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reason for this revision surgery was reported as dislocation.The previous surgery and the surgery detailed in this event occurred 21 days apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the product that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within their expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to dislocation.There were no findings during this evaluation that indicate that the reported device was defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may contribute to an event that are outside of the control of djo surgical such as poor bone density, inadequate soft tissue support, excessive range of motion, patient activities or trauma.Due to the short time between previous and revision surgery, it is also possible that the event may have been occurred due to improper implant selection, lack of post-operative care.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Event or Problem Description
Revision surgery - patients hip dislocated.Removed and exchanged out the bipolar and head.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPOLAR HIP
Common Device Name
SHELL, BIPOLAR, MODULAR, ENCORE, SIZE45
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
MDR Report Key9081377
Report Number1644408-2019-00876
Device Sequence Number2700502
Product Code KWY
UDI-Device Identifier00888912023436
UDI-Public(01)00888912023436
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K953510
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/10/2022
Device Model Number412-02-045
Device Catalogue Number412-02-045
Device Lot Number990B1173
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/19/2019
Initial Report FDA Received Date09/18/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age66 YR
-
-