• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON PRIM CEM FXD BPLT #2; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON PRIM CEM FXD BPLT #2; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 5520B200
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Codes Unspecified Infection (1930); Injury (2348); Inadequate Osseointegration (2646)
Date of Event 07/16/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
An event regarding revision due to failed right total knee arthroplasty was reported.Clinician review of the provided medical records indicated incomplete cementation of the tibial baseplate as well as late onset hematogenous infection of the arthroplasty and arthroscopic debridement of the knee may have increased ligament laxity of the arthroplasty knee.The reported loosening and instability were confirmed however infection was not confirmed.Method & results: product evaluation and results: not performed as no product was returned for evaluation.Clinician review: a review of the provided medical records by a clinical consultant indicated: suboptimal initial device fixation through incomplete cementation of the tibial baseplate has contributed to loosening requiring revision surgery.It would be quite likely that a late onset low- grade infection of hematogenous origin has contributed as well to failure while prior to revision an extensive arthroscopic debridement of the knee may have contributed to increased ligament laxity thereby increasing complaints requiring more early intervention with revision surgery of all components, replaced with a competitor's device with an adequate end result.Product history review: indicate all devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot or sterile lot referenced.The following devices were also listed in this report: triathlon symmetric x3 patella; cat#5550-g-278; lot#z563, triathlon cr fem comp #2 r-cem; cat#5510f202; lot#scfaj, it cannot be determined which, if any of these devices may have caused or contributed to the patient's experience.Conclusion: a review of the provided medical records by a clinical consultant indicated: suboptimal initial device fixation through incomplete cementation of the tibial baseplate has contributed to loosening requiring revision surgery.It would be quite likely that a late onset low-grade infection of hematogenous origin has contributed as well to failure while prior to revision an extensive arthroscopic debridement of the knee may have contributed to increased ligament laxity thereby increasing complaints requiring more early intervention with revision surgery of all components, replaced with a competitor's device with an adequate end result.No further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.The information in this report was provided by stryker orthopaedics legal affairs department.No additional information is available at this time due to the ongoing litigation.Should additional information become available, the evaluation summary will be submitted in a supplemental report.
 
Event or Problem Description
It was reported through the communication of an attorney that allegedly the patient was revised due to "failed right total knee arthroplasty".Update (b)(6) 2019 by of, as per medical review: suboptimal initial device fixation through incomplete cementation of the tibial baseplate has contributed to loosening requiring revision surgery.It would be quite likely that a late onset low- grade infection of hematogenous origin has contributed as well to failure while prior to revision an extensive arthroscopic debridement of the knee may have contributed to increased ligament laxity thereby increasing complaints requiring more early intervention with revision surgery of all components, replaced with a competitor's device with an adequate end result.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON PRIM CEM FXD BPLT #2
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key9081721
Report Number0002249697-2019-03242
Device Sequence Number19553988
Product Code MBH
UDI-Device Identifier07613327041569
UDI-Public07613327041569
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K141056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2007
Device Explanted Year2016
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number5520B200
Device Lot NumberSBX9J
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/22/2019
Initial Report FDA Received Date09/18/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age81 YR
-
-