DEPUY ORTHOPAEDICS INC US PIN LNR CONS NEUT +4 36IDX56OD; PINNACLE HIP SYSTEM : HIP CONSTRAINED ACETABULAR LINERS Back to Search Results
Catalog Number 121836656
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/21/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Per internal procedures, the event information was reviewed.No investigational inputs were received.For this investigation, no immediate action was required as no alleged deficiency with the device was identified.No device associated with this report was received for examination.In order to determine if a lot related issue was possible, a worldwide complaint database search was performed.A worldwide complaint database search found no additional related reports against the provided product code/lot number combination.Based on the inability to find any additional related reports against the provided product code/lot number combination, it is reasonable to conclude that there are no anomalies with regard to manufacturing or inspection contained in the device history records that would contribute to the reported event.Investigational inputs were requested as indicated per internal procedures for this failure mode, however, no information was provided to depuy.Without the physical complaint sample associated with this report, it was not possible to determine if the device failed to meet specifications at the time it was released for distribution.The device associated with this event was used in the treatment of the patient as prescribed by the presiding surgeon.From the event information received, it was not possible to determine the relationship of the device to the reported event.No evidence was found indicating product error was a contributing factor.The need for corrective action was not indicated.Depuy considers the investigation closed.Should additional info be received, the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that after revision the surgeon's plan was to convert to a constrained liner and ts ceramic head.Removed primary liner and head.Implanted constrained liner and put on 36+12 ts ceramic head.While doctor was reducing the hip he crushed the side of the constrained liner.The surgeon then removed the liner and head again.And then implanted a new constrained liner and a ts ceramic 36+8.5 head.After reducing the patient they ran him through a range of motion where the liner was pulled from the cup.They took everything back out and put it back in.They ran him through a range of motion again and the liner disassociated from the cup again.At that point we took everything out and implanted a 456x40 +4 10 degree liner with a 40 bipolar head.After running everything through a range of motion doctor was happy with stability and he closed up.Doe: (b)(6) 2019.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PIN LNR CONS NEUT +4 36IDX56OD
Common Device Name
PINNACLE HIP SYSTEM : HIP CONSTRAINED ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9081843
Report Number1818910-2019-105187
Device Sequence Number19488767
Product Code KWZ
UDI-Device Identifier10603295012122
UDI-Public10603295012122
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K043058
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number121836656
Device Lot NumberH19890
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/21/2019
Initial Report FDA Received Date09/18/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age61 YR
-
-