| Catalog Number |
750310 |
| Medical Device Problem Code |
Component Missing (2306)
|
| Health Effect - Clinical Codes |
No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that there was no piston syringe included in the tray.
|
| |
|
Event or Problem Description
|
|
It was reported that there was no piston syringe included in the tray.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed.The product could not be used for treatment since only the packaging was included.The device did not meet specifications, which was influenced by the reported failure.Visual evaluation of the sample noted one opened piston irrigation syringe tray.Upon looking into the tray, it was noted that the tray was empty aside from the tray packaging.Missing components were not permitted.Although the reported event was confirmed, the root cause could not be determined.A potential root cause for this failure could be machine speed out of parameters.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: ¿contents: 1 graduated solution container 1 graduated plastic collection basin.1 50 cc (1-2/3 oz) bulb syringe".
|
| |
|