• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® URINE METER WITH BAG; DRAIN BAG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® URINE METER WITH BAG; DRAIN BAG Back to Search Results
Model Number 153202
Medical Device Problem Code Restricted Flow rate (1248)
Health Effect - Clinical Codes Abdominal Pain (1685); Incontinence (1928); Urinary Retention (2119); Patient Problem/Medical Problem (2688)
Date of Event 08/23/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the patient allegedly experienced abdominal pain with full bladder sensation due to no urine flow.The patient was probed in the intensive care unit.Urine would not flow, just drip.A control bladder scan was performed on the patient and 900 cc was found in the patient's bladder.
 
Additional Manufacturer Narrative
The device was not returned for evaluation.A potential failure mode could be 'does not allow for urine drainage¿ with a potential root cause of 'vent blocked creating vacuum in bag, poor interfaces with connections, occlusions.¿ the device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: ¿visually inspect the product for any imperfections or surface deterioration prior to use.If package is opened or if any imperfection or surface deterioration is observed, do not use.1.Remove protective cap from drainage tube catheter adapter and connect drainage tube to catheter.2.Position hanger on bedside rail, using string or hook.3.Use sheeting clip to secure drainage tube to sheet.Important: hang drainage tube in a straight fashion from bedside to drainage bag.Ensure that the drainage bag is placed near the foot of the bed.4.If using a urine meter, it may be emptied in two ways: a.To empty into the bag, grasp the bottom of the meter and lift up.To ensure that the meter empties completely, lifting again is recommended.B.To empty urine meter into receptacle, twist green portion of drain valve to the left; to close, twist green position of the drain valve to the right.5.To empty bag: a.Remove outlet tube from housing; gently squeeze connector arms and pull tube from housing.B.Release clamp and empty bag.C.After emptying, reclamp outlet tube and slide connector into housing until connector arms engage.Note: if specimen is required, see directions for using urine sample port.6.Periodic observations of this system should be made to ensure that urine is flowing freely.If a standing column of urine is observed, check for correct positioning of bag and then for a physical obstruction.If correct positioning or removal of physical obstruction does not allow free flow, the bag may have to be changed.7.If bag is not positioned correctly, urine may bypass the meter and go directly to the bag.Reposition bag as necessary.Directions for using bard ez-lok® sampling port: bard ez-lok® sampling port accepts a luer-lock or slip tip syringe.1.Kink drainage tubing a minimum of 3 inches below the sampling port until urine is visible under the access site.2.Swab surface of site with antiseptic wipe.3.Using aseptic technique, position the syringe in the center of the sampling port.The syringe should be held perpendicular to the surface of the sampling port (at approximately 80-100 degree angle).Press the syringe firmly and twist gently to lock the syringe onto the sampling port.Note: improper penetration technique could cause formation of a drop of urine on the surface of the sampling port.Periodic observation of the sampling port is recommended.4.Aspirate desired volume of urine.5.Unkink tubing and send specimen to laboratory." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Event or Problem Description
It was reported that the patient allegedly experienced abdominal pain with full bladder sensation due to no urine flow.The patient was probed in the intensive care unit.Urine would not flow, just drip.A control bladder scan was performed on the patient and 900 cc was found in the patient's bladder.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® URINE METER WITH BAG
Common Device Name
DRAIN BAG
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key9082863
Report Number1018233-2019-05742
Device Sequence Number2022779
Product Code KNX
UDI-Device Identifier00801741033100
UDI-Public(01)00801741033100
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date (Section B) 09/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2023
Device Model Number153202
Device Catalogue Number153204
Device Lot NumberNGCZ3308
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 08/27/2019
Supplement Date Received by Manufacturer09/25/2019
Initial Report FDA Received Date09/18/2019
Supplement Report FDA Received Date09/27/2019
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/31/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-