| Model Number |
0502-444-070 |
| Medical Device Problem Code |
Optical Problem (3001)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/22/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device manufacture date is not known at this time.However, should it become available it will be provided in future reports.Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported image was foggy during procedure.Additional information confirmed a replacement was used to complete the procedure successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was not returned for investigation therefore the reported failure mode was not confirmed.The failure mode will be monitored for future reoccurrence .The device manufacture date is not known.Alleged failure: foggy.Probable root cause: ¿ laser welding seal failure ¿ distal/proximal window solder failure ¿ damage to optical train ¿ damage to needle ¿ damage to distal or proximal windows ¿ moisture intrusion ¿ end of life wear-out ¿ environmental disturbance: endoscope colder than dew point ¿ environmental disturbance: fluid entrapment between the coupler and eyepiece junction (eyepiece only) ¿ use error.
|
| |
|
Event or Problem Description
|
|
It was reported image was foggy during procedure.Additional information confirmed a replacement was used to complete the procedure successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: foggy probable root cause: ¿ laser welding seal failure ¿ distal/proximal window solder failure ¿ damage to optical train ¿ damage to needle ¿ damage to distal or proximal windows ¿ moisture intrusion ¿ end of life wear-out ¿ environmental disturbance: endoscope colder than dew point ¿ environmental disturbance: fluid entrapment between the coupler and eyepiece junction (eyepiece only) ¿ use error the product was not returned for investigation therefore the reported failure mode was not confirmed.The failure mode will be monitored for future reoccurrence .
|
| |
|
Event or Problem Description
|
|
Per additional information received, the fogging could be wiped off.The same device could be used to complete the procedure.The foggy camera head and scope may cause user inconvenience but did not result in any adverse consequences.Therefore making this event not reportable.
|
| |
|