MAUDE Adverse Event Report: PENUMBRA, INC. RUBY COIL; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. RUBY COIL; HCG, KRD Back to Search Results
Catalog Number RBY4C1035
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/09/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has been returned and the investigation results are pending.A follow up mdr will be submitted upon completion of the device investigation.This report is associated with mfr report numbers: 3005168196-2019-01775, 3005168196-2019-01776.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the internal iliac artery (iia) using ruby coils.During the procedure, the physician successfully placed a pod coil.Upon advancement of a ruby coil into the target location, the physician decided to retract the coil to reposition it, but reported that it unintentionally detached in the target location.The same event occurred with the next ruby coil.However, both ruby coils were implanted successfully and therefore remained in the patient.While checking a third ruby coil on the back table, the physician found that it had detached.Therefore, the ruby coil was not used in the procedure.The procedure was completed using five additional ruby coils and six other coils.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Results: the pet lock was intact on the proximal end of the pusher assembly.The pull wire was not within the pusher assembly distal detachment tip (ddt) alignment feature.The embolization coil was undamaged and detached from the pusher assembly.The introducer sheath was ovalized approximately 36.5 cm from the proximal end.Conclusions: evaluation of the returned ruby coil revealed the pull wire was not within the pusher assembly ddt alignment feature.If the ruby coil is mishandled during use, the pusher assembly may elongate past the reach of the pull wire.If this occurs, the embolization coil will likely detach from the pusher assembly.Further evaluation revealed an ovalization on the introducer sheath.This damage was likely incidental to the reported complaint.Penumbra coils are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RUBY COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key9082988
Report Number3005168196-2019-01777
Device Sequence Number2837651
Product Code HCG
UDI-Device Identifier00814548018669
UDI-Public00814548018669
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
K173614
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/21/2019
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberRBY4C1035
Device Lot NumberF87018
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/06/2019
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 08/21/2019
Supplement Date Received by Manufacturer09/25/2019
Initial Report FDA Received Date09/18/2019
Supplement Report FDA Received Date10/18/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age71 YR
Date Report Sent to FDA01/01/2005
-
-