The customer reported that the multiple patient receiver (org-9110a) failed and displayed an "error light" message, causing comm loss at the central nurse's station (cns) and the remote network station (rns).The cns and rns were monitoring 8 tele beds.No consequence or impact to the patients were reported.No consequence or impact to the patients were reported.Nihon kohden sent the customer an exchange org to resolve the issue.The defective org was returned to nihon kohden and is awaiting evaluation.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.Additional model information: the following devices were being used in conjunction with the org: cns, rns, tele devices.No model and serial information is known as attempts to obtain information were made, but not provided.
|
Details of complaint: the customer reported that the multiple patient receiver (org-9110a) failed, displaying an "error light" message, causing comm loss at the central nurse's station (cns) and the remote network station (rns).The cns and rns were monitoring 8 tele beds.No patient harm was reported.Nihon kohen technical support (nkts) advised the customer to power cycle the org, but the issue persisted.The caller stated they were not trained to reconfigure the org, requesting to send the unit in for an exchange.An exchange unit was shipped to the customer on 09/10/2019.No patient harm was reported.Service requested / performed: exchange / evaluation: no physical damage was noticed.Upon repair center (rc) evaluation, the reported problem was duplicated.An "invalid crc-32 value in rom" error message was displayed.On 05/22/2020, the org was reprogrammed to sw ver 03-31, which resolved the issue.The unit was tested per the operator's/service manual and the results were recorded on the maintenance check sheet.The unit completed 48 hours of extended testing and operated to the manufacturer's specifications.Investigation summary: the possible causes for a crc error to appear when using a computer, are corruption of a file or program registry, file and setting configuration mistakes, and/or faulty app or program installations.A problem with the hard drive can also cause a cyclic redundancy check data error.The overall risk of this event is determined to be medium.The reported issue does not require further investigation through capa process since the issue is known to be caused due to factors outside of nk device capabilities.Additional device information: d10 concomitant medical device: the following devices were used in conjunction with the org: cns: model #: ni; serial #: ni.Rns: model #: ni; serial #: ni.Transmitters: model #: ni; serial #: ni.
|