| Catalog Number |
07D70-21 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/26/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.All available patient information was included.Additional patient details are not available.Complete patient identifier: (b)(6).
|
| |
|
Event or Problem Description
|
|
The customer reported a falsely elevated magnesium result generated on the architect c8000 analyzer on one patient.Results provided: sid (b)(6), (b)(6) 2019 = 6.49 / 2.17 / 2.05 mg/dl; new tube= 2.01 mg/dl.Customers normal range: (1.8-2.6 mg/dl).No impact to patient management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
It was discovered on october 18, 2019 that the initial emdr for this issue was submitted under the correct suspect medical device but incorrect manufacturer name, city and state.Mdr number 1628664-2019-00674 has been submitted for the correct manufacturing site and all further information will be documented under that mdr number.
|
| |
|