| Model Number |
TTA2 |
| Medical Device Problem Code |
Patient Data Problem (3197)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/29/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A thorough evaluation of this event will be included in a follow-up report upon investigation completion.
|
| |
|
Event or Problem Description
|
|
A customer reported measurement data can appear from a previous study onto a new study if an error message appears during launch while using the tomtec-arena tta 2.This event occurred during a research study.There was no injury or clinical use associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
This action was reported to fda per 21 cfr part 806 on 10/2/2019.Reference corrections and removal report number (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
This action was reported to fda per 21 cfr part 806.Reference corrections and removal report number 8043836-10/02/19-001-c.
|
| |
|
Additional Manufacturer Narrative
|
|
A thorough technical investigation was performed which included evaluation of workflow details, a software design issue was determined to be to cause and a solution has been included in a newly released software revision.
|
| |
|
Search Alerts/Recalls
|