MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. EP-WORKMATE¿ SCU-SIGNAL CONDITIONING UNIT (AMPLIFIER), 56 CHANNEL; PROGRAMMABLE DIAGNOSTIC COMPUTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. EP-WORKMATE¿ SCU-SIGNAL CONDITIONING UNIT (AMPLIFIER), 56 CHANNEL; PROGRAMMABLE DIAGNOSTIC COMPUTER Back to Search Results
Model Number 09-1462-0056
Medical Device Problem Code Failure to Analyze Signal (1539)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/17/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The results, method and conclusion codes along with investigation results will be provided in the final report.
 
Event or Problem Description
During pacing there was no signal displayed on the live screen when the catheter was connected to the pin box.All connections were checked and cabling was replaced, but the issue remained.The procedure was cancelled and there were no adverse consequences to the patient.
 
Event or Problem Description
During pacing there was no signal displayed on the live screen when the catheter was connected to the pin box.All connections were checked, and cabling was replaced, but the issue remained.The procedure was cancelled and there were no adverse consequences to the patient.During further troubleshooting, it was found that the system reference cable going into the amplifier was loose.Once that was tightened, the issue was resolved.
 
Additional Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The review determined the process was performed and completed in accordance with abbott specifications and procedures.Based on the information received, the cause of the reported incident was due to a loose cable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EP-WORKMATE¿ SCU-SIGNAL CONDITIONING UNIT (AMPLIFIER), 56 CHANNEL
Common Device Name
PROGRAMMABLE DIAGNOSTIC COMPUTER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
MDR Report Key9086651
Report Number2184149-2019-00172
Device Sequence Number2537762
Product Code DQK
UDI-Device Identifier05415067003126
UDI-Public05415067003126
Combination Product (Y/N)N
Initial Reporter CountryHK
PMA/510(K) Number
K151911
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number09-1462-0056
Device Lot Number4960831
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 09/17/2019
Supplement Date Received by Manufacturer09/19/2019
Initial Report FDA Received Date09/19/2019
Supplement Report FDA Received Date10/08/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-