Two broken instruments were tagged from sterile processing department.The 32mm x 64n blue liner trial was used over the weekend for a cup extraction.Zimmer cup extraction instrument was used and scratched/ damaged out edge of the cup trial.Sz 8 sz 7 20mm attune shim has a small crack in it and was handed to me from sterile processing.Was surgery delayed due to the reported event? : unknown, was procedure successfully completed? : yes, were fragments generated? : yes, if yes, were they removed easily without additional intervention? : yes, patient status/ outcome / consequences : no, was other medical intervention (e.G.X-rays, additional procedures, prescriptions, otc, revision) required: unknown, is the patient part of a clinical study : unknown, (b)(4).Device property of :none, device in possession of :none, (b)(4).Device property of :none, device in possession of :none.By checking this box i certify that all information that are known/available has been disclosed.If any new information will be made available, the additional information will be submitted through cst.True.
|
|
Depuy synthes is submitting this report pursuant to the provision depuy synthes of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: corrected: h3 (device evaluated by mfr) and h6.Product complaint#: (b)(4).Investigation summary: investigation of the returned device revealed the shim has cracked.The device is not broken as initially reported.The investigation findings with this individual complaint did not indicate that a broader investigation or corrective action was necessary.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot: null.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
|