The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample analysis, applicable fmea documents, applicable manufacturing records, and labeling.Based on a review of this information, the following was concluded: the complaint of a missing needle cover is inconclusive due to poor sample condition.One 19 x 0.75 in powerloc infusion set was returned for investigation with opened packaging.The lot number on the label showed ascps0013.The white dust caps are present on both luer hubs.The safety mechanism was not activated over the needle bevel.No evidence of use was observed throughout the device.The needle cover was not returned.No punctures or damage to the packaging was observed.The presence of the needle cover prior to package opening could not be determined from the sample provided; therefore, the complaint is inconclusive due to poor sample condition.A lot history review (lhr) of ascps0013 showed no other similar product complaint(s) from this lot number.
|