• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. KIT MMT-1755KR 630G 3ML V4.10 EN RECERT; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. KIT MMT-1755KR 630G 3ML V4.10 EN RECERT; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-1755KR
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/13/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
The customer reported via phone call that the insulin pump had insulin pump error.Customer¿s blood glucose level was 206 mg/dl.Customer was able to clear the alarm and was not able to complete rewind.Customer was advised to place insulin pump in storage mode.The customer was also advised that the insulin pump needed to be replaced and was advised to discontinue use of the insulin pump and to revert to the backup plan.The device will be returned for analysis.
 
Additional Manufacturer Narrative
Device pass self test, active current measurement test, and sleep current measurement test.Device received error 39 during rewind due to corroded motor home switch and high magnetic field exposure.P-cap or reservoir will not lock properly due to missing retainer.Unable to perform displacement test, prime or seating test, basic occlusion test, force sensor test, and occlusion test due to pump error 39.
 
Additional Manufacturer Narrative
This report is part of a retrospective review and remediation efforts in response to a warning letter.Updated h9: z-0955-2020.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KIT MMT-1755KR 630G 3ML V4.10 EN RECERT
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
8185464805
MDR Report Key9096148
Report Number2032227-2019-65880
Device Sequence Number16874740
Product Code OZO
UDI-Device Identifier00763000179632
UDI-Public(01)00763000179632(17)211226
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date12/26/2021
Device Model NumberMMT-1755KR
Device Catalogue NumberMMT-1755KR
Device Lot NumberHG32F8K
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/01/2019
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 09/13/2019
Supplement Date Received by Manufacturer11/15/2019
10/14/2019
Initial Report FDA Received Date09/19/2019
Supplement Report FDA Received Date11/16/2019
05/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/27/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0955-2020
Patient Sequence Number1
Patient Age65 YR
Patient SexFemale
Patient Weight222 KG
-
-