| Model Number |
MMT-723RNAP |
| Medical Device Problem Codes |
Connection Problem (2900); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
07/26/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the reservoir compartment was damaged.The customer¿s blood glucose level was 300 mg/dl.Troubleshooting was performed for damage and noticed the reservoir did not lock in place when inserted into insulin pump.Customer was advised to the insulin pump would need to be replaced, to discontinue use of the insulin pump and revert to a back-up plan.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
A test reservoir was installed and did not lock in place due to completely broken reservoir tube lip.Device had missing reservoir tube o-ring.
|
| |
|