| Model Number |
MMT-722LNAP |
| Medical Device Problem Codes |
Battery Problem (2885); Power Problem (3010); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/13/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump had unspecified error.The customer¿s blood glucose was unknown at the time of incident.The customer also stated that the insulin pump had diminished battery life and also rejected new battery.The insulin pump will not be returned for analysis.
|
| |
|
Search Alerts/Recalls
|