MAUDE Adverse Event Report: MEDTRONIC MINIMED SENSOR ENLITE MMT-7008A; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MINIMED SENSOR ENLITE MMT-7008A; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-7008A
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Rash (2033); Blood Loss (2597)
Event Date 09/16/2019
Event Type  malfunction  
Manufacturer Narrative
Inspected one opened/used sensor and performed continuity resistance test, sensor passed per specification.Then performed a bicarbonate buffer test, and sensor passed per specifications with accurate readings.
 
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
 
Event Description
It was reported that the customer had site issue due to sensor.Customer's blood glucose level was 7.0 mmol/l.Customer reported that they received medical treatment for site issue.Customer did see the blood on sensor after insertion.Sensor had already been removed from body.Customer applied some rubbing alcohol.The sensor will not be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSOR ENLITE MMT-7008A
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC MINIMED
18000 devonshire st.
northridge CA 91325 1219
Manufacturer (Section G)
MEDTRONIC MINIMED
18000 devonshire st.
northridge CA 91325 1219
Manufacturer Contact
gerwin de graaff
18000 devonshire st.
northridge, CA 91325-1219
8185464805
MDR Report Key9096691
MDR Text Key183083083
Report Number2032227-2019-66212
Device Sequence Number1
Product Code OZO
Combination Product (y/n)N
Reporter Country CodeCA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberMMT-7008A
Device Catalogue NumberMMT-7008A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/06/2020
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/16/2019
Initial Date FDA Received09/20/2019
Supplement Dates Manufacturer Received03/09/2020
Supplement Dates FDA Received03/12/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
-
-