| Model Number |
MMT-722NAP |
| Medical Device Problem Code |
Device Difficult to Program or Calibrate (1496)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/11/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device received with intermittent button response due to corroded keypad traces.No button error alarm during testing.Device passed self test.The asr exemption e2015043 was revoked on february 4, 2019.This event was previously reported as an asr.Additional information or analysis results regarding this event will be reported as an mdr.
|
| |
|
Event or Problem Description
|
|
It was reported that the insulin pump alarmed button error and the buttons does not work.Customer¿s blood glucose was unknown at the time of incident.No troubleshooting was performed.Insulin pump is expected to return for analysis.
|
| |
|