| Catalog Number |
0038470 |
| Medical Device Problem Code |
Misassembled (1398)
|
| Health Effect - Clinical Codes |
No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/26/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that the cap on the tip was missing.
|
| |
|
Event or Problem Description
|
|
It was reported that the cap on the tip was missing.Per additional information received from the investigator on 26sep2019, the cap is a removable piece that is placed in the syringe tip.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed.Visual evaluation of the sample noted one opened unused piston syringe.No cap(cm1880) was noted to be on the syringe tip, but an adaptor was included within the syringe barrel s intended.Missing components are not allowed per procedure.Although the reported event was confirmed, the root cause could not be determined.A potential root cause for this failure could be error of inspector.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "caution: secure tip tightly prior to use." correction: d4.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Search Alerts/Recalls
|