| Catalog Number |
5MAXJET7KIT |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Vasoconstriction (2126)
|
| Date of Event |
06/19/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The product was not returned for evaluation.From the information provided, there is no indication that there was any device malfunction, non conformance, or misuse that contributed to the reported event.Potential adverse events in the labeling with the penumbra system include, but are not limited to, vessel spasm, hematoma or hemorrhage at the site, inability to completely remove thrombus, intracranial hemorrhage, ischemia, including death.Therefore, it was determined that the reported adverse event was anticipated complication.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 3005168196-2019-01404.
|
| |
|
Event or Problem Description
|
|
The patient was undergoing a thrombectomy procedure in the left middle cerebral artery (mca) using a penumbra system 3d revascularization device (psr3d) and a penumbra system jet7 kit (kit).During the procedure, while making an initial pass using the psr3d with the jet7 and a non-penumbra microcatheter, the physician experienced resistance while attempting to insert the psr3d into a non-penumbra microcatheter and, therefore, the psr3d was retracted back into the delivery sheath.The physician then attempted to insert the same psr3d, but the same issue occurred.Therefore, the physician removed the psr3d and found the leading edge of the distal tip to be kinked.The procedure was completed using a new psr3d with the same jet7 and microcatheter.After the procedure, moderate vasospasm was present in the distal internal carotid artery.It was also reported that the vasospasm occurred upon placing the jet7.The vasospasm was considered resolved as of (b)(6) 2019.
|
| |
|
Event or Problem Description
|
|
The patient was undergoing a thrombectomy procedure in the left middle cerebral artery (mca) using a penumbra system 3d revascularization device (psr3d), a penumbra system jet7 kit (kit) and, a neuron max 6f 088 long sheath (neuron max).During the procedure, while making an initial pass using the psr3d with the jet7, neuron max and, a non-penumbra microcatheter, the physician experienced resistance while attempting to insert the psr3d into a non-penumbra microcatheter and, therefore, the psr3d was retracted back into the delivery sheath.The physician then attempted to insert the same psr3d, but the same issue occurred.Therefore, the physician removed the psr3d and found the leading edge of the distal tip to be kinked.The procedure was completed using a new psr3d with the same jet7, neuron max and microcatheter.After the procedure, moderate vasospasm was present in the distal left internal carotid artery.It was also reported that the vasospasm occurred upon placing the neuron max.The vasospasm was considered resolved as of (b)(6) 2019.
|
| |
|
Additional Manufacturer Narrative
|
|
Please note that the following sections are being updated based on additional information provided by a penumbra representative on 10/16/2019: 1.Section b.Box 5.Describe event or problem.Please note that the following section was incorrectly reported on the initial mfr report and is being corrected on this follow-up #1 3005168196-2019-01797 mfr report: 1.Section b.Box 5.Describe event or problem.This report is associated with mfr report numbers: 1.3005168196-2019-01404; 2.3005168196-2019-02104.
|
| |
|