| Model Number |
W3DR01 |
| Medical Device Problem Codes |
Reset Problem (3019); Data Problem (3196)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/21/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the implantable pulse generator (ipg) underwent a diagnostic reset and consequently battery voltage data was unavailable.The device was reprogrammed and remains in use.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the device was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated a power on reset occurred.Analysis of the device memory indicated that the atrial and the ventricular rate histogram data was missing/invalid.Analysis of the device memory indicated the battery measurement was not available.If information is provided in the future, a supplemental report will be issued.
|
| |
|