MEDTRONIC HEART VALVES DIVISION CG FUTURE BAND; RING, ANNULOPLASTY Back to Search Results
Model Number 638RL26
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Mitral Regurgitation (1964)
Date of Event 09/12/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product analysis: no product was returned.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information that immediately post implant of this 26mm mitral annuloplasty ring, it was explanted and replaced with a non-medtronic bioprosthetic valve.The reason for replacement was reported as moderate residual regurgitation.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
Product analysis: upon receipt at medtronic¿s quality laboratory, visual evaluation showed damage on the sewing cloth adjacent to the trigonal marker, likely resulting from suture placement through the eyelet.From the inflow, the sewing cloth and green suture was frayed.Radiography showed no evidence of distortion or fractures in the ring.Conclusion: surgeons often attempt to repair valves in lieu of replacing them in an effort to preserve the native mitral anatomy.There are times when valve repair is attempted using a repair device and subsequent post-repair evaluation demonstrates an inadequate result.This is potentially due to, but not limited to, sub-optimal anatomy, surgical technique or inappropriate sizing, and not a malfunction of the device.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CG FUTURE BAND
Common Device Name
RING, ANNULOPLASTY
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
MDR Report Key9104421
Report Number2025587-2019-02890
Device Sequence Number2165040
Product Code KRH
UDI-Device Identifier00643169793514
UDI-Public00643169793514
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K061127
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/12/2023
Device Model Number638RL26
Device Catalogue Number638RL26
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/11/2019
Initial Date Received by Manufacturer 09/12/2019
Supplement Date Received by Manufacturer01/15/2020
Initial Report FDA Received Date09/23/2019
Supplement Report FDA Received Date02/11/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-