| Model Number |
638RL26 |
| Medical Device Problem Codes |
Adverse Event Without Identified Device or Use Problem (2993); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Mitral Regurgitation (1964)
|
| Date of Event |
09/12/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product analysis: no product was returned.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that immediately post implant of this 26mm mitral annuloplasty ring, it was explanted and replaced with a non-medtronic bioprosthetic valve.The reason for replacement was reported as moderate residual regurgitation.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis: upon receipt at medtronic¿s quality laboratory, visual evaluation showed damage on the sewing cloth adjacent to the trigonal marker, likely resulting from suture placement through the eyelet.From the inflow, the sewing cloth and green suture was frayed.Radiography showed no evidence of distortion or fractures in the ring.Conclusion: surgeons often attempt to repair valves in lieu of replacing them in an effort to preserve the native mitral anatomy.There are times when valve repair is attempted using a repair device and subsequent post-repair evaluation demonstrates an inadequate result.This is potentially due to, but not limited to, sub-optimal anatomy, surgical technique or inappropriate sizing, and not a malfunction of the device.If information is provided in the future, a supplemental report will be issued.
|
| |
|