| Model Number |
TRM-255 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.This is a known potential adverse event addressed in the product labeling.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported rupture and "textured claim".Device was explanted.This record is for the left side.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported rupture and "textured claim".Device was explanted.This record is for the left side.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified an opening and red particles.Weight measurement of the device identified device was underweight.A microscopic analysis was performed which identified a striated edge opening consistent with use of a surgical tool.Based on the device analysis the final assessment was a striated edge opening on radius side due to surgical damage.
|
| |
|